Auto-Tuning Distortion Meter
HM-250



 

Outline
The HM-250 is a distortion meter capable of 0.1% full scale measurement continuously over the frequency range of 20 Hz to 20 kHz??A???Ewith a typical residual distortion of 0.01%. It makes full use of pushbutton switches and provides a calibration which can be activated with a single operation regardless of the selected distortion measurement range. Carrier frequency elimination is performed by an automatic tuning technique??A???Ethereby greatly simplifying operation. The HM-250 is also usable as a high-sensitivity (100?…?EV f.s.) AC voltmeter.
Features
Automatic Tuning
To eliminate the signal carrier component, simply perform a rough adjustment. The automatic tuning function thereafter automatically tunes to the carrier for extremely simple operation.

High Sensitivity
The HM-250 provides 0.1% full scale as a distortion meter, and 100μV full-scale measurements when used as an AC voltmeter 
(10 Hz to 200 kHz).

Single-Operation Calibration
Calibration can be activated with a single operation regardless of the selected distortion measurement range.

X-Y Outputs for Waveform Observation
When measuring distortion, the input signal carrier component is provided as the X-axis output and the harmonic component is provided as the Y-axis output (while the input signal is output in output voltage measurements).



SPECIFICATIONS
SPECIFICATIONS
Distortion measurement
Measurement range 0.1% to 100% full scale? in 1-3 steps (7 ranges)
Input level 100mV to 100V (for 100% CAL)30mV min. (automatic tuning is possible)
Carrier rejection ratio 80dB min.
Basic frequency range 20Hz to 20kHz in three ranges? and fine adjustment
Residual distortion 0.01% max. (typ)? (1kHz? Sensitivity max.)
Harmonic characteristics 20Hz to 10kHz carrier: ±1dB10Hz to 20kHz carrier: ±2dB
(for a harmonic frequency of 60kHz or lower and distortion of 30% or less)
AC voltage measurement
Measurement range 100μV to 100V full scale? in 1-3 steps
Indication accuracy ±3% of full scale max. (referenced to 1kHz)
Residual noise 5% of full scale indication or lower (input shorted? 0.1mV range)
Frequency response 20Hz to 20kHz: ±0.5dB
10Hz to 200kHz: ±1dB
(referenced 1kHz)
Common specifications
Indication Mean-value detection with rms-value indication (calibrated for a sinewave)
X output Approx. 1Vrms (tuned at 100% CAL) (Output of input signal carrier component)
Y output Approx. 0.5Vrms (at full scale)
Harmonic component output for distortion measurements? and input signal output for output voltage measurement
X and Y outputs impedance Approx. 2.2kΩ (unbalanced)
Input impedance Approx. 100kΩ? 100pF max. (unbalanced)
Maximum input voltage At ATT 0dB:
 120Vrms (60Hz or less)
 200VDC or 200V (DC ACpeak)
At ATT 30dB or 60dB:
 140Vrms
 200VDC or 200V (DC ACpeak)
Power source 100/120/220/240V (max. 250V) AC? 50/60Hz? Approx. 5W
Case dimensions 212 (W)×133 (H)×272 (D) mm
Maximum dimensions 212 (W) ×156 (H)×288 (D) mm
Weight Approx. 3.7kg
Accessories Instruction manual (1) 
Power cord (1) 
Accessory cable (2)