RF INPUT |
|
Frequency Range |
20㎒ ~ 2.5㎓ |
|
|
|
Level Range |
-30dBm ~ + 20dBm |
|
|
|
Modulation Frequency |
25㎐ to 50㎑ |
|
|
|
Input Impedance |
50Ω |
|
|
|
FM MEASUREMENT |
|
Range Selection |
100, 50, 30, 10, FM3 |
|
|
|
Accuracy |
±(2%of FS+2% reading +residual FM ) |
|
|
|
Sensitivity |
20~ 1500 MHz : -30 ~ 20 dBm 1500MHz ~ 2500MHz : -20 ~ 20 dBm |
|
|
|
Distortion |
< 1.5%, for 50kHz dev. at 1kHz Mod.freq |
|
|
|
De-Emphasis |
OFF, 50, 75, 750㎲ ( Time constant accuracy =±5% ) |
|
|
|
Filter |
LPF : OFF, 3kHz, 15kHz, 50kHz HPF : OFF, 25Hz, 300kHz, 3kHz |
|
|
|
AM MEASUREMENT |
|
Range Selection |
100, 50, 30, 10 |
|
|
|
Accuracy |
±(3% of FS + 2% of reading + residual AM ) |
|
|
|
Sensitivity |
20~1000MHz : -30~0 dBm |
|
|
|
Residual AM |
< 3% for 30% depth at 1kHz Mod.freq |
|
|
|
SIGNAL OUTPUT |
|
AF Output |
Connector : Front panel BNC Frequency : 25㎐ to 50㎑ Level : 0.5Vrms Impedance : 600Ω |
|
|
|
Distortion |
FM : <1%(for 7㎑ deviation at 1㎑ Mod-Freq) AM : < 3%(for 30% AM at 1㎑ Mod-Freq) |
|
|
|
RF LEVEL MEASUREMENT |
|
RF Input |
Frequency : 20㎒ to 2.5㎓ Input Level : -30dBm to 20dBm Input Impedance : 50Ω |
|
|
|
Accuracy |
±1dB |
|
|
|
Display |
7 segment 4 digits |
|
|
|
Unit |
dBm, dBu, mW |
|
|
|
FREQUENCY COUNTER |
|
Frequency |
20㎒ to 2.5㎓ |
|
|
|
Gate Time |
0.1sec to 1sec(0.1, 0.3, 0.5, 1s) |
|
|
|
Digit |
7 segment 9 digits Display |
|
|
|
Reading Speed |
9digit/sec |
|
|
|
Time Base |
OCXO (5×10-8) |
|
|
|
OTHERS |
|
Power Requirements |
115/230 VAC , 50/60㎐, Auto Selectable |
|
|
|
Dimensions |
115(H) × 430(W) × 310(D) mm | |